Dissemin is shutting down on January 1st, 2025

Published in

Royal Society of Chemistry, Journal of Analytical Atomic Spectrometry, 7(26), p. 1526

DOI: 10.1039/c0ja00271b

Links

Tools

Export citation

Search in Google Scholar

P, S and Cl trace detection by laser ablation double-focusing sector field ICP-MS to identify local defects in coated glasses

This paper is available in a repository.
This paper is available in a repository.

Full text: Download

Green circle
Preprint: archiving allowed
Orange circle
Postprint: archiving restricted
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

Glass defects are always undesired because of their significant costs to manufacturing industries. Laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS) is a versatile technique for local trace element analysis in a wide variety of solid samples. Furthermore, the coupling of laser systems to double-focusing sector field mass analyzers provides low limits of detection with a high mass resolving power, allowing the accurate analysis of P, S and Cl traces in local defects of coated glasses. Three different types of defects, depending on the impurities identified at the LA-ICP-MS profiles, were found in the selected coated glasses (with metallic and oxide films in the low nanometre range), enabling the unambiguous identification of the contamination source that produces the local defects.