Published in

Cambridge University Press, High Power Laser Science and Engineering, (5), 2017

DOI: 10.1017/hpl.2017.10

Links

Tools

Export citation

Search in Google Scholar

An automated, 0.5 Hz nano-foil target positioning system for intense laser plasma experiments

This paper is made freely available by the publisher.
This paper is made freely available by the publisher.

Full text: Download

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Green circle
Published version: archiving allowed
Data provided by SHERPA/RoMEO

Abstract

We report on a target system supporting automated positioning of nano-targets with a precision resolution of $4~\unicode[STIX]{x03BC}\text{m}$ in three dimensions. It relies on a confocal distance sensor and a microscope. The system has been commissioned to position nanometer targets with 1 Hz repetition rate. Integrating our prototype into the table-top ATLAS 300 TW-laser system at the Laboratory for Extreme Photonics in Garching, we demonstrate the operation of a 0.5 Hz laser-driven proton source with a shot-to-shot variation of the maximum energy about 27% for a level of confidence of 0.95. The reason of laser shooting experiments operated at 0.5 Hz rather than 1 Hz is because the synchronization between the nano-foil target positioning system and the laser trigger needs to improve.