Published in

Cambridge University Press, Radiocarbon, 2(22), p. 461-469, 1980

DOI: 10.1017/s0033822200009772

Links

Tools

Export citation

Search in Google Scholar

Low-Level Gas Proportional Counting in an Underground Laboratory

Journal article published in 1980 by Hh H. Loosli, Martin Heimann ORCID, Hans Oeschger
This paper is made freely available by the publisher.
This paper is made freely available by the publisher.

Full text: Download

Green circle
Preprint: archiving allowed
Red circle
Postprint: archiving forbidden
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

Results of measurements in an underground laboratory, 70m of water equivalent below surface, are given. Background values of proportional gas counters with volumes between 16cc and 1.5L are lower by a factor of 2—4.5 compared to the values in the previously used laboratory (7m w.e.). High counting gas pressures, up to 5 at CH4 and up to 36 at P-10, enable the use of relatively small counters with correspondingly small background contributions from the walls. A separation of the residual background into different components is attempted, distinguishing a pressure dependent volume effect and surface correlated contributions. It can be shown that the selection of radioactively pure counter construction material is very important for a good low level counting system.