Circulating free light chain measurement in the diagnosis, prognostic assessment and evaluation of response of AL amyloidosis: comparison of Freelite and N latex FLC assays

Distributing this paper is prohibited by the publisher
Distributing this paper is prohibited by the publisher

Full text: Unavailable

Red circle
Preprint: archiving forbidden
Red circle
Postprint: archiving forbidden
Orange circle
Published version: archiving restricted
Data provided by SHERPA/RoMEO

Abstract

AbstractBackground:The measurement of circulating free light chain (FLC) is essential in the diagnosis, prognostic stratification and evaluation of response to therapy in light chain (AL) amyloidosis. For more than 10 years, this has been done with an immunonephelometric assay based on polyclonal antibodies (Freelite), and cutoffs for staging and response assessment have been validated with this method. Recently, a new assay based on monoclonal antibodies (N latex FLC) has been marketed in Europe.Methods:We evaluated and compared the clinical performance of the two assays in 426 patients with newly diagnosed AL amyloidosis.Results:We found suboptimal agreement between the two methods, with differences between values obtained with the Freelite and N latex FLC assays increasing with the concentration of clonal FLC. The diagnostic sensitivity of the Freelite (82%) and N latex FLC (84%) assays was similar, and both improved to 98% in combination with serum and urine immunofixation. The concentration of FLC measured with both methods had prognostic significance. Less pronounced decreases in FLC best predicted improved survival with the N latex FLC assay (33% vs. 50%), and there was poor concordance (84%) in discrimination of responders.Conclusions:The two assays have similar diagnostic and prognostic performance. However, they are not interchangeable, and follow-up should be done with either one. New response criteria are needed for the N latex FLC assay.