Published in

Elsevier, Journal of Virological Methods, 1-2(186), p. 137-140

DOI: 10.1016/j.jviromet.2012.07.023

Links

Tools

Export citation

Search in Google Scholar

Comparison of Xpert Flu rapid nucleic acid testing with rapid antigen testing for the diagnosis of influenza A and B

This paper is available in a repository.
This paper is available in a repository.

Full text: Download

Green circle
Preprint: archiving allowed
Red circle
Postprint: archiving forbidden
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

Influenza infections are associated with thousands of hospital admissions and deaths each year. Rapid detection of influenza is important for prompt initiation of antiviral therapy and appropriate patient triage. In this study the Cepheid Xpert Flu assay was compared with two rapid antigen tests, BinaxNOW Influenza A & B and BD Directigen EZ Flu A + B, as well as direct fluorescent antibody testing for the rapid detection of influenza A and B. Using real-time, hydrolysis probe-based, reverse transcriptase PCR as the reference method, influenza A sensitivity was 97.3% for Xpert Flu, 95.9% for direct fluorescent antibody testing, 62.2% for BinaxNOW, and 71.6% for BD Directigen. Influenza B sensitivity was 100% for Xpert Flu and direct fluorescent antibody testing, 54.5% for BinaxNOW, and 48.5% for BD Directigen. Specificity for influenza A was 100% for Xpert Flu, BinaxNOW, and BD Directigen, and 99.2% for direct fluorescent antibody testing. All methods demonstrated 100% specificity for influenza B. These findings support the use of the Xpert Flu assay in settings requiring urgent diagnosis of influenza A and B.