A four-dimensional photon detector for PET application

This paper was not found in any repository, but could be made available legally by the author.
This paper was not found in any repository, but could be made available legally by the author.

Full text: Unavailable

Green circle
Preprint: archiving allowed
Orange circle
Postprint: archiving restricted
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

We analyzed a photon detector for positron emission tomography with high spatial resolution and depth of interaction capability. The detector is composed of a monolithic LYSO scintillator crystal coupled on top and bottom sides to two custom SiPM arrays. We investigated the ability to reconstruct the DOI of the 511 keV photon comparing the number of triggered SiPMs on the two sides of the module. Acquisitions were performed scanning the lateral surface of the crystal with a collimated 511 keV photon beam at different incident positions. A standard deviation of 1.5 mm in depth of interaction was obtained at the center of the module.