32 channels SPAD array for single photon timing applications

Proceedings article published in 2012 by C. Scarcella, S. Bellisai, A. Della Frera, A. Ruggeri ORCID, S. Tisa, A. Tosi, F. Zappa
This paper was not found in any repository, but could be made available legally by the author.
This paper was not found in any repository, but could be made available legally by the author.

Full text: Unavailable

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

Many applications require photodetectors with single-photon sensitivity and high timing resolution. We present a module based on a linear array of 32 Single-Photon Avalanche Diodes (SPADs) for multi-channel Time-Correlated Single-Photon Counting (TCSPC) applications. Each pixel includes a 30 μm active area diameter SPAD and quenching and counting electronics. We measured a timing resolution of about 110 ps, a photon detection efficiency of 40% at λ = 420 nm and a very low dark count rate of about 150 cps at room temperature.