Published in

SAGE Publications, Magnetic Resonance Insights, (5), p. MRI.S9832, 2012

DOI: 10.4137/mri.s9832

Links

Tools

Export citation

Search in Google Scholar

IR-TrueFISP: A New DCE-MRI Approach in Comparison with 3D-Flash Multi Flip Angle Method

Journal article published in 2012 by Buechert ORCID, Martin Büchert, Klaus Mross
This paper is made freely available by the publisher.
This paper is made freely available by the publisher.

Full text: Download

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

For assessing treatment response to novel cancer therapeutics, dynamic contrast enhanced magnetic resonance imaging (DCE-MRI) is a valuable tool. Quantitative functional parameter estimates can be obtained by fitting physiological models to the data. In this study the IR-TrueFISP approach was evaluated as DCE-MRI acquisition sequence against the widely used 3D-Flash protocol. For comparison both protocols providing different spatial and temporal information were investigated in phantom and patient examinations. 12 advanced tumor patients underwent two examinations on consecutive days using both protocols. Results were compared and were in good agreement with each other. IR-TrueFISP data showed a lower variability compared to 3D-Flash results. This work demonstrates the pros and cons of both investigated methods. It was demonstrated that the known IR-TrueFISP sequence can successfully be employed as DCE-MRI acquisition method estimating perfusion parameters. The benefits of the IR-TrueFISP protocol are high temporal resolution and good accuracy.