A 9.4 T Fourier transform ion cyclotron resonance mass spectrometer: description and performance

Distributing this paper is prohibited by the publisher
Distributing this paper is prohibited by the publisher

Full text: Unavailable

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

9.4 Tesla Fourier transform ion cyclotron resonance (FT-ICR) mass spectrometers (Bruker BioAPEX-94e) have been installed at the Division of Ion Physics, Uppsala University, and at the Department of Chemistry, University of Warwick. The BioAPEX-94e FT-ICR instrument is built around a high-field, superconducting magnet and a platform with easily interchangeable ion sources [matrix-assisted laser desorption/ionisation (MALDI), secondary ion mass spectrometry (SIMS), electrospray ionisation (ESI) and electron impact/chemical ionisation (EI/CI)]. In this paper a technical description of the instrument is given. Outstanding performance characteristics are demonstrated, notably clear resolution of C59N+ and C5813C2+ (mass difference 3.65 mDa) and mass measurement accuracy at the low ppm level. A wide range of applications in Warwick and Uppsala is described, demonstrating the versatility and high performance of the instrument.