Automated Blood Pressure Determination During Exercise Test. Clinical Evaluation of a New Automated Device

This paper was not found in any repository, but could be made available legally by the author.
This paper was not found in any repository, but could be made available legally by the author.

Full text: Unavailable

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

The accuracy and reproducibility of a new automatic device (P) specially designed for noninvasive blood pressure monitoring during the exercise stress test were evaluated in 50 consecutive subjects (34 normotensives and 16 hyper tensives). Automatic measurements were compared with those taken by a sphygmomanometer (RR). A good agreement between systolic pressure values obtained by the two methods was found (RR 159±30 mmHg, P 158 ± 28 mmHg, mean difference = -1.53 ± 13 mmHg, p = 0. 166, ns). On the contrary the new device significantly underestimated diastolic pressure values (RR 89.3 ± 13 mmHg; P 84±13 mmHg, mean difference -5.37±9.3, p < 0.001). In conclusion the new device seems able to measure systolic but underesti mates diastolic blood pressure both in hypertensives and in normotensives dur ing the effort test.