Links

Tools

Export citation

Search in Google Scholar

Thermal effects of cement mantle thickness for hip resurfacing.

Journal article published in 2008 by Jp Paige Little, J. Paige Little, Ha Gray, Dw Murray, Dj Beard, Hs Gill ORCID
This paper is available in a repository.
This paper is available in a repository.

Full text: Download

Question mark in circle
Preprint: policy unknown
Question mark in circle
Postprint: policy unknown
Question mark in circle
Published version: policy unknown

Abstract

Hybrid hip resurfacing arthroplasty with uncemented acetabular and cemented femoral fixation is increasingly becoming popular as an alternative to total hip arthroplasty. There is concern about femoral neck fractures, and long-term survival has not yet been demonstrated. Thermal necrosis may be an important factor for neck fracture and will affect the viability of the femoral bone. This cadaveric study investigated the thermal effect of thick (1.5 mm, n = 3) and thin (0.5 mm, n = 3) cement mantles; 5 thermocouples were used to record temperature at the femoral bone/cement interface during hip resurfacing arthroplasty. The highest recorded temperatures were significantly higher when a thick cement mantle is used (45.4 degrees C), compared to a thin cement mantle (32.7 degrees C). To reduce the potential for thermal necrosis, the thin cement mantle technique is recommended.