Published in

Elsevier, Thrombosis Research: Vascular Obstruction, Hemorrhage and Hemostasis, 3(130), p. 535-537, 2012

DOI: 10.1016/j.thromres.2012.05.028

Links

Tools

Export citation

Search in Google Scholar

Biological variation of INR in stable patients on long-term anticoagulation with warfarin

This paper was not found in any repository, but could be made available legally by the author.
This paper was not found in any repository, but could be made available legally by the author.

Full text: Unavailable

Green circle
Preprint: archiving allowed
Orange circle
Postprint: archiving restricted
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

Within-individual biological variation of INR (CV(B)) was assessed in 245 selected stable warfarin-treated patients monitored by three thrombosis centers. Selection criteria were: treatment period of six months or longer before the observation period; at least six consecutive INRs within the therapeutic range of 2.0 - 3.0; interval between consecutive INR measurements of two weeks or longer; no change in warfarin dose; no changes in the patient's circumstances which may influence the INR, such as intercurrent diseases, invasive procedures, starting or stopping drugs interacting with warfarin. The minimum, maximum and mean within-individual coefficient of variation CV(B) of the INR measurements in the 245 selected patients were 0.4%, 14.5%, and 9.0%, respectively Analytical performance goals for the INR measurement (imprecision) could be derived from the mean CV(B). For a therapeutic range of 2.0 - 3.0 with warfarin, the desirable and optimum imprecision of INR determination is