Dissemin is shutting down on January 1st, 2025

Published in

Korean Academy of Medical Sciences, Journal of Korean Medical Science, 10(25), p. 1462, 2010

DOI: 10.3346/jkms.2010.25.10.1462

Links

Tools

Export citation

Search in Google Scholar

Long-term Prognosis of Left Ventricular Lead

This paper is made freely available by the publisher.
This paper is made freely available by the publisher.

Full text: Download

Red circle
Preprint: archiving forbidden
Green circle
Postprint: archiving allowed
Green circle
Published version: archiving allowed
Data provided by SHERPA/RoMEO

Abstract

Transvenous left ventricular (LV) lead implantation is on the increase due to cardiac resynchronization therapy (CRT). However, there has been paucity of data on the prognosis of LV lead. Consecutive 32 patients with LV lead for CRT (n=22) or pacemaker (n=10) were subjected. Serial changes in pacing threshold and impedance along with lead-related complications were evaluated. Over 2 yr follow-up, there was no significant change in relative threshold voltage to the initial value (100%, 110%, 89.6%, and 79.6% at baseline, 1, 6, and 24 months respectively, P=0.62) as well as lead impedance (816±272, 650±178, 647±191, and 590±185 ohm at baseline, 1, 6, and 24 months respectively, P=0.80). The threshold change was not affected by lead position, lead polarity, and indication of lead implantation. The cumulative rates of lead revision were 6.3% (n=2) and 9.4% (n=3) in 6 month and 2 yr follow-up, respectively. One case of phrenic nerve capture at left lateral decubitus position was detected 1 month after the implantation. However, there were no serious complications over 2 yr period. In conclusion, transvenous LV lead implantation showed favorable long-term prognosis. Pacing parameters remained stable without significant changes over 2 yr follow-up.