Evaluation of a next generation direct whole blood enzymatic assay for hemoglobin A1c on the ARCHITECT c8000 chemistry system

Distributing this paper is prohibited by the publisher
Distributing this paper is prohibited by the publisher

Full text: Unavailable

Red circle
Preprint: archiving forbidden
Red circle
Postprint: archiving forbidden
Orange circle
Published version: archiving restricted
Data provided by SHERPA/RoMEO

Abstract

AbstractThe utility of HbAHere we completed a method evaluation of the ARCHITECT HbAThe total imprecision ranged from 0.5% to 2.2% CV with low and high level control materials. Around the diagnostic cut-off of 48 mmol/mol, the total imprecision was 0.6% CV. Mean bias using reference samples from IFCC and CAP ranged from –1.1 to 1.0 mmol/mol. The enzymatic assay also showed excellent agreement with HPLC methods, with slopes of 1.01 and correlation coefficients ranging from 0.984 to 0.996 compared to Menarini Adams HA-8160, Bio-Rad Variant II and Variant II Turbo instruments. Finally, no significant effect was observed for erythrocyte sedimentation or interference from common hemoglobin variants in patient samples containing heterozygous HbS, HbC, HbD, HbE, and up to 10% HbF.The ARCHITECT enzymatic assay for HbA