Objective. To evaluate the relationship between spectral-domain optical coherence tomography (OCT) and standard automated perimetry (SAP) in healthy and glaucoma individuals. Methods. The sample comprised 338 individuals divided into 2 groups according to intraocular pressure and visual field outcomes. All participants underwent a reliable SAP and imaging of the optic nerve head with the Cirrus OCT. Pearson correlations were calculated between threshold sensitivity values of SAP (converted to linear scale) and OCT parameters. Results. Mean age did not differ between the control and glaucoma groups (59.55 ± 9.7 years and 61.05 ± 9.4 years, resp.; P = 0.15). Significant differences were found for the threshold sensitivities at each of the 52 points evaluated with SAP (P