Published in

2015 37th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC)

DOI: 10.1109/embc.2015.7319486

Links

Tools

Export citation

Search in Google Scholar

Robot Assisted Stapedotomy ex vivo with an Active Handheld Instrument

This paper is available in a repository.
This paper is available in a repository.

Full text: Download

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

Micron is a fully handheld active micromanipulator that helps to improve position accuracy and precision in microsurgery by cancelling hand tremor. This work describes adaptation, tuning, and testing of the Micron system for stapedotomy, a microsurgical procedure performed in the middle ear to restore hearing that requires accurate manipulation in narrow spaces. Two end-effectors, a handle, and a brace (or rest) were designed and prototyped. The control system was adapted for the new hardware. The system was tested ex vivo in stapedotomy procedure comparing manually-performed and Micron-assisted surgical tasks. Tremor amplitude was found to be reduced significantly. Further testing is needed in order to obtain statistically significant results regarding other parameters dealing with regularity of the fenestra shape