Operational performance characteristics of the WISH detector array on the ISIS spallation neutron source

Journal article published in 2014 by D. Duxbury, D. Khalyavin ORCID, P. Manuel ORCID, D. Raspino, N. Rhodes, E. Schooneveld, E. Spill
This paper is made freely available by the publisher.
This paper is made freely available by the publisher.

Full text: Download

Green circle
Preprint: archiving allowed
Orange circle
Postprint: archiving restricted
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

The performance of the position sensitive neutron detector array of the WISH diffractometer is discussed. WISH (Wide angle In a Single Histogram) is one of the seven instruments currently available for users on the second target station (TS2) of the ISIS spallation neutron source, and is used mainly for magnetic studies of materials. WISH is instrumented with an array of 10 detector panels, covering an angular range of 320o, orientated in two semi-cylindrical annuli around a central sample position at a radius of 2.2m. In total the 10 detector panels are composed of 1520 3He based position sensitive detector tubes. Each tube has an active length of one metre, a diameter of 8mm and is filled with 3He at 15 bar. The specification for the WISH detectors included a neutron detection efficiency of 50% at a neutron wavelength of 1Å with good gamma rejection. A position resolution better than 8 mm FWHM along the length of the tubes was also required which has been met experimentally. Results obtained from the detector arrays showing pulse height and positional information both prior to and post installation are shown. The first 5 of the 10 detector panels have been operational since 2009, and comparable diffraction data from powder and single crystal samples taken from the remaining 5 panels (installation completed in 2013) shows that we have a detector array with a highly stable performance which is easily assembled and maintained. Finally some real user data is shown, highlighting the excellent quality of data attainable with this instrument.