Validation of 3DVH estimated DVH metrics for prostate VMAT plans

Journal article published in 2015 by Sankar Arumugam, Aitang Xing, Tony Young, David Thwaites, Lois Holloway ORCID
This paper is made freely available by the publisher.
This paper is made freely available by the publisher.

Full text: Download

Red circle
Preprint: archiving forbidden
Red circle
Postprint: archiving forbidden
Green circle
Published version: archiving allowed
Data provided by SHERPA/RoMEO

Abstract

The accuracy of 3DVH (Sun Nuclear Corporation, USA) reported DVH metrics for target volumes and Organs at Risk (OARs) for two Prostate Volumetric Modulated Arc Therapy (VMAT) plans was studied. The accuracy of 3DVH estimated DVH metrics in the presence of Multi Leaf Collimator (MLC) systematic errors was also tested with error introduced plans calculated in Pinnacle. The results of the study show that the DVH metrics estimated by 3DVH for error-free plans agree with the TPS calculation within 3%. The D95 to PTV was shown to be sensitive in detecting studied MLC errors. However the accuracy of 3DVH estimated DVH metrics for Target Volumes and OARs in the presence of MLC errors for VMAT prostate plans has limitations with this small data set. Although for most situations values matched within 3% for small MLC errors, there was up to a 9.8% difference between the TPS and 3DVH in the presence of a simulated 5mm MLC positioning error. Further study with more plans including other treatment sites is required to fully assess the performance of 3DVH in detecting potential clinical delivery errors.