Published in

Wiley, Electroanalysis, 13(17), p. 1198-1206, 2005

DOI: 10.1002/elan.200503237

Links

Tools

Export citation

Search in Google Scholar

Understanding Capacitively Coupled Contactless Conductivity Detection in Capillary and Microchip Electrophoresis. Part 1. Fundamentals

This paper is available in a repository.
This paper is available in a repository.

Full text: Download

Green circle
Preprint: archiving allowed
Orange circle
Postprint: archiving restricted
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

Capacitively coupled contactless conductivity detection (C4D) is presented in a progressively detailed approach. Through different levels of theoretical and practical complexity, several aspects related to this kind of detection are addressed, which should be helpful to understand the results as well as to design a detector or plan experiments. Simulations and experimental results suggest that sensitivity depends on: 1) the electrolyte co-ion and counter-ion; 2) cell geometry and its positioning; 3) operating frequency. Undesirable stray capacitance formed due to the close placement of the electrodes is of great importance to the optimization of the operating frequency and must be minimized.