Published in

Elsevier, Spectrochimica Acta Part B: Atomic Spectroscopy, 2(57), p. 339-353

DOI: 10.1016/s0584-8547(01)00398-6

Links

Tools

Export citation

Search in Google Scholar

A procedure for correcting self-absorption in calibration free-laser induced breakdown spectroscopy

This paper is available in a repository.
This paper is available in a repository.

Full text: Download

Green circle
Preprint: archiving allowed
Red circle
Postprint: archiving forbidden
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

A model of the self-absorption effect in laser-induced plasma has been developed, with the aim of providing a tool for its automatic correction in the Calibration-Free algorithm recently developed for standardless analysis of materials by LIBS (Laser Induced Breakdown Spectroscopy). As a test of the model, the algorithm for self-absorption correction is applied to three different certified steel NIST samples and to three ternary alloys (Au, Ag, Cu) of known composition. The experimental results show that the self-absorption corrected Calibration-Free method gives reliable results, improving the precision and the accuracy of the CF-LIBS procedure by approximately one order of magnitude.