New setup for angular distribution measurements of Auger electrons from fixed in space molecules

This paper is available in a repository.
This paper is available in a repository.

Full text: Download

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Orange circle
Published version: archiving restricted
Data provided by SHERPA/RoMEO

Abstract

A new experimental setup for measurement of the angular distributions of energy selected Auger electrons emitted from fixed in space molecules is presented. The system is based on two identical ion detectors with a small angular acceptance placed at 0 degrees and 90 degrees relative to the polarization axis of the incident radiation, and a high luminosity double-toroidal electron analyzer combined with position sensitive detection. This setup allows selection of the molecular alignment for sigma and pi ionization channels simultaneously and provides an energy and ejection angle measurement of the outgoing electron. The performance in terms of energy and angular resolution, as well as the calibration procedure, are discussed. Sample results obtained on the carbon monoxide ionized above the C 1s threshold are presented. (C) 2000 American Institute of Physics. [S0034- 6748(00)01801-9].