Dissemin is shutting down on January 1st, 2025

Published in

13th IEEE International Conference on BioInformatics and BioEngineering

DOI: 10.1109/bibe.2013.6701537

Links

Tools

Export citation

Search in Google Scholar

Modeling stent deployment in realistic arterial segment geometries: The effect of the plaque composition

This paper is available in a repository.
This paper is available in a repository.

Full text: Download

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

Stents are medical devices used in cardiovascular intervention for unblocking the diseased arteries and restoring blood flow. During stent implantation the deformation of the arterial wall as well as the resulted stresses caused in the arterial morphology are studied. In this paper we study the effect of the composition of the atherosclerotic plaque during the stent deployment procedure, using Finite Element modeling. The stenting procedure is simulated for two different cases; in the first the presence of the plaque is ignored whereas in the second a three dimensional (3D) stiff calcified plaque is located in the stenotic area of the artery. Results indicate that in the second case the von Mises stresses in the arterial wall are higher than the stresses occurred in the first case. In addition, the distribution of the arterial von Mises stress depends on the plaque composition.