Compact soft x-ray multichord camera: Design and initial operation

This paper is available in a repository.
This paper is available in a repository.

Full text: Download

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Orange circle
Published version: archiving restricted
Data provided by SHERPA/RoMEO

Abstract

A compact and low cost diagnostic for spatially resolved measurements of soft x-ray or total radiation emission has been designed and realized to be flexibly applied to different plasma physics experiments. Its reduced size (outer diameter=35 mm) makes it suited to a variety of devices. The line integrated emissivity (brightness) has been measured along up to 20 lines of sight, using an array of miniaturized silicon photodiodes. Preliminary prototypes of the diagnostic have been installed in the Madison Symmetric Torus reversed field pinch (RFP) device at University of Wisconsin and in the EXTRAP T2 RFP device at the Royal Institute of Technology in Stockholm. Application of the diagnostic to a gas-fed (argon, helium) magnetoplasma dynamic thruster (MPDT) with an external magnetic field will also be discussed. © 2003 American Institute of Physics.