Published in

Nature Research, Nature Methods, 11(4), p. 915-918, 2007

DOI: 10.1038/nmeth1108

Links

Tools

Export citation

Search in Google Scholar

STED microscopy with continuous wave beams

Journal article published in 2007 by Katrin I. Willig, Benjamin Harke, Rebecca Medda, Stefan W. Hell
This paper is made freely available by the publisher.
This paper is made freely available by the publisher.

Full text: Download

Green circle
Preprint: archiving allowed
Orange circle
Postprint: archiving restricted
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

We report stimulated emission depletion (STED) fluorescence microscopy with continuous wave (CW) laser beams. Lateral fluorescence confinement from the scanning focal spot delivered a resolution of 29-60 nm in the focal plane, corresponding to a 5-8-fold improvement over the diffraction barrier. Axial spot confinement increased the axial resolution by 3.5-fold. We observed three-dimensional (3D) subdiffraction resolution in 3D image stacks. Viable for fluorophores with low triplet yield, the use of CW light sources greatly simplifies the implementation of this concept of far-field fluorescence nanoscopy.