Determination of coupled solvent mass in quartz crystal microbalance measurements using deuterated solvents

Journal article published in 2003 by Vincent S. J. Craig ORCID, Mark Plunkett
This paper is available in a repository.
This paper is available in a repository.

Full text: Download

Green circle
Preprint: archiving allowed
Red circle
Postprint: archiving forbidden
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

A simple method is described for determining of the contribution of hydrodynamically coupled solvent to the adsorbed film mass determined in a quartz crystal microbalance (QCM) when operated in liquid. The method requires no additional apparatus and utilizes the change in QCM resonant frequency response between measurements made in non-deuterated and deuterated solvents. The mass of coupled water in a polymer film has been determined and is found to agree with that determined by XPS analysis of the dried polymer film.