Validation of F-Scan pressure sensor system: A technical note

Journal article published in 1998 by Z. P. Luo, L. J. Berglund, K. N. An
This paper is available in a repository.
This paper is available in a repository.

Full text: Download

Question mark in circle
Preprint: policy unknown
Question mark in circle
Postprint: policy unknown
Question mark in circle
Published version: policy unknown

Abstract

This study performed a quantitative validation on a recently developed pressure sensitive transducer system, the F-Scan sensor system. The results indicate that the sensor is adequate for determination of pressure distribution under contact conditions with soft materials. The linear response of the sensor was up to 1.7 MPa with good homogeneity throughout sensor cells. However, the sensor is sensitive to surface conditions, loading speeds, and temperature. Variations also exist from sensor to sensor. In order to have accurate measurement, calibration was recommended in actual clinical or experimental conditions prior to use, including surface contact conditions, loading speeds, and temperature environment. In addition, this sensor system is not suitable for hard surface contact such as plexiglas.