Published in

Elsevier, Biosensors and Bioelectronics, 9(18), p. 1095-1100

DOI: 10.1016/s0956-5663(02)00245-2

Links

Tools

Export citation

Search in Google Scholar

An amperometric bi-enzyme sensor for determination of formate using cofactor regeneration

This paper is available in a repository.
This paper is available in a repository.

Full text: Download

Green circle
Preprint: archiving allowed
Red circle
Postprint: archiving forbidden
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

A biosensor for detection of formate at submicromolar concentrations has been developed by co-immobilizing formate dehydrogenase (FDH, E.C. 1.2.1.2), salicylate hydroxylase (SHL, E.C. 1.14.13.1) and NAD(+) linked to polyethylene glycol (PEG-NAD(+)) in a poly(vinyl alcohol) (PVA) matrix in front of a Clark-electrode. The principle of the bi-enzyme scheme is as follows: formate dehydrogenase converts formate into carbon dioxide using PEG-NAD(+). Corresponding PEG-NADH produced is then oxidized to PEG-NAD(+) by salicylate hydroxylase using sodium salicylate and oxygen. The oxygen consumption is monitored with the Clark-electrode. The advantages of this biosensor approach are the effective re-oxidation of PEG-NADH, and the entrapment of PEG-NAD(+) resulting in avoiding the addition of expensive cofactor to the working medium for each measurement. This bi-enzyme sensor has achieved a linear range of 1-300 microM and a detection limit of 1.98 x 10(-7) M for formate (S/N=3), with the response time of 4 min. The working stability is limited to 7 days due to the inactivation of the enzymes. Only sodium salicylate was needed in milli-molar amounts.