Published in

Elsevier, Journal of Environmental Sciences, 5(25), p. 895-907

DOI: 10.1016/s1001-0742(12)60251-4

Links

Tools

Export citation

Search in Google Scholar

An online monitoring system for atmospheric nitrous acid (HONO) based on stripping coil and ion chromatography

This paper is available in a repository.
This paper is available in a repository.

Full text: Download

Green circle
Preprint: archiving allowed
Red circle
Postprint: archiving forbidden
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

A new instrument for measuring atmospheric nitrous acid (HONO) was developed, consisting of a double-wall glass stripping coil sampler coupled with ion chromatography (SC-IC). SC-IC is featured by small size (50 x 35 x 25 cm) and modular construction, including three independent parts: the sampling unit, the transfer and supporting unit, and the detection unit. High collection efficiency (> 99%) was achieved with 25 micromol/L Na2CO3 as absorption solution even in the presence of highly acidic compounds. This instrument has a detection limit of 8 pptv at 15 min time resolution, with a measurement uncertainty of 7%. Potential interferences from NO(x), NO2+SO2, NO2+VOCs, HONO+O3, HNO3, peroxyacetyl nitrite (PAN) and particle nitrite were quantified in laboratory studies and were found to be insignificant under typical atmospheric conditions. Within the framework of the 3C-STAR project, inter-comparison between the SC-IC and LOPAP (long path liquid absorption photometer) was conducted at a rural site in the Pearl River Delta. Good agreement was achieved between the two instruments over three weeks. Both instruments determined a clear diurnal profile of ambient HONO concentrations from 0.1 to 2.5 ppbv. However, deviations were found for low ambient HONO concentrations (i.e. < 0.3 ppbv), which cannot be explained by previous investigated interference species. To accurately determine the HONO budget under illuminated conditions, more intercomparison of HONO measurement techniques is still needed in future studies, especially at low HONO concentrations.