Published in

Wiley, Electroanalysis, 12(27), p. 2792-2799, 2015

DOI: 10.1002/elan.201500236

Links

Tools

Export citation

Search in Google Scholar

A highly sensitive sensor based on reduced graphene oxide, carbon nanotube and a Co(II) complex modified carbon paste electrode: Simultaneous determination of isoprenaline, captopril and tryptophan

This paper is available in a repository.
This paper is available in a repository.

Full text: Download

Green circle
Preprint: archiving allowed
Orange circle
Postprint: archiving restricted
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

This paper describes the development of a reduced graphene oxide (RGO), carbon nanotube (CNT) and Co(II) complex (cobalt(II) bis (benzoylacetone) ethylenediimino) (CBE) modified carbon paste electrode (CPE) for simultaneous determination of isoprenaline (IP), captopril (CAP) and tryptophan (Try). A pair of well-defined redox peaks of Co(II) complex were obtained through a direct electron transfer between the Co(II) complex and the CPE. The proposed sensor showed very efficient electrocatalytic activity for anodic oxidation of IP in a 0.1 M phosphate buffer solution (pH 7.0). Square wave voltammetry (SWV) exhibited two linear dynamic ranges of 0.125–30.0 µM and 30.0–300.0 µM for IP. The detection limit for IP was found to be 50 nM. The proposed sensor was successfully applied for the determination of IP in real samples such as human blood serum, urine and IP ampoule.