Reduction of image artifacts in three-dimensional optical coherence tomography of skin in vivo

Journal article published in 2011 by Yih Miin Liew ORCID, Robert A. McLaughlin, Fiona M. Wood, David D. Sampson ORCID
This paper is available in a repository.
This paper is available in a repository.

Full text: Download

Red circle
Preprint: archiving forbidden
Green circle
Postprint: archiving allowed
Green circle
Published version: archiving allowed
Data provided by SHERPA/RoMEO

Abstract

This paper presents results of in vivo studies on the effect of refractive index-matching media on image artifacts in optical coherence tomography (OCT) images of human skin. These artifacts present as streaks of artificially low backscatter and displacement or distortion of features. They are primarily caused by refraction and scattering of the OCT light beam at the skin surface. The impact of the application of glycerol and ultrasound gel is assessed on both novel skin-mimicking phantoms and in vivo human skin, including assessment of the epidermal thickening caused by the media. Based on our findings, recommendations are given for optimal OCT imaging of skin in vivo.