Oxford University Press, Interactive Cardiovascular and Thoracic Surgery, 2(9), p. 150-154
DOI: 10.1510/icvts.2008.201418
Full text: Download
We developed a new color charge-coupled device (CCD) camera for the intraoperative indocyanine green (ICG) angiography. This device consists of a combination of custom-made optical filters and an ultra-high sensitive CCD image sensor, which can detect simultaneously color and near-infrared (NIR) rays from 380 to 1200 nm. We showed a comparison between our system and other devices for the preliminary experience. We routinely performed both transit-time flowmetry (TFM) and color images for intraoperative assessment, thallium-scintigraphy for the early postoperative assessment, and then angiography after 1-year surgery. We also obtained intraoperative graft flows and images in 116 grafts. Although TFM indicated a graft patency, the CCD camera suspected perfusion failures in four grafts. Also the analysis of the ICG fluorescence intensity showed the significant hypoperfusion at the perfusion territory distal to the anastomosis (graft vs. perfusion territory; 230+/-26 vs. 156+/-13 a.u, P=0.02). When the CCD camera suspected a graft failure, CCD camera and angiography showed a comparable graft failure. The unique device that visualized ICG-enhanced structures against a background of natural myocardial color improved the visibility of abnormality in flow and perfusion. Our findings show that this device may become a standard intraoperative graft and perfusion assessment tool in coronary artery bypass graft (CABG).