Evaluation of the Abbott m 2000 RealTi m e Human Immunodeficiency Virus Type 1 (HIV-1) Assay for HIV Load Monitoring in South Africa Compared to the Roche Cobas AmpliPrep-Cobas Amplicor, Roche Cobas AmpliPrep-Cobas TaqMan HIV-1, and BioMerieux NucliSENS EasyQ HIV-1 Assays

This paper was not found in any repository, but could be made available legally by the author.
This paper was not found in any repository, but could be made available legally by the author.

Full text: Unavailable

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Orange circle
Published version: archiving restricted
Data provided by SHERPA/RoMEO

Abstract

ABSTRACT The implementation of antiretroviral therapy demands the need for increased access to viral load (VL) monitoring. Newer real-time VL testing technologies are faster and have larger dynamic ranges and fully automated extraction to benefit higher throughputs in resource-poor environments. The Abbott RealTi m e human immunodeficiency virus type 1 (HIV-1) assay was evaluated as a new option for testing for HIV-1 subtype C in South Africa, and its performance was compared to the performance of existing assays (the Cobas AmpliPrep-Cobas TaqMan HIV-1, version 1, assay; the AmpliPrep-Cobas Monitor standard HIV-1 assay; and the NucliSENS EasyQ-EasyMag HIV-1 assay) in a high-throughput laboratory. The total precision of the RealTi m e HIV-1 assay was acceptable over all viral load ranges. This assay compared most favorably with the Cobas AmpliPrep-Cobas TaqMan HIV-1 assay ( R 2 = 0.904), with a low standard deviation of difference being detected (0.323 copies/ml). The bias against comparator assays ranged from −0.001 copies/ml to −0.228 copies/ml. Variability in the reporting of VLs for a 20-member subtype panel compared to the variability of other assays was noted with subtypes G and CRF02-AG. The RealTi m e HIV-1 assay can test 93 samples per day with minimal manual preparation, less staff, and the minimization of contamination through automation. This assay is suitable for HIV-1 subtype C VL quantification in South Africa.