An Advanced Preamplifier for Highly Segmented Germanium Detectors

Journal article published in 2006 by A. Pullia ORCID, F. Zocca, G. Pascovici
This paper is available in a repository.
This paper is available in a repository.

Full text: Download

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

We present a fast low-noise hybrid charge-sensitive preamplifier for germanium position-sensitive gamma-ray detectors. In conjunction with a bulky 36-fold segmented detector it provided an excellent resolution of 1.71/1.77 keV fwhm on the 1.17/1.33MeV60Co lines. The preamplifier rise time, as measured at the test bench, is as fast as 7.5 ns, with a detector capacitance of 21 pF and with a 5 m 50 Omega twisted-pair cable connected at its output. The dynamic range of the preamplifier input stage is as large as 92 dB, ranging from 0.275 fC to 9.9 pC, i.e., from 5 keV to 180 MeV in terms of photon energy. On signals larger than 2 MeV a fast reset (~ 10 MeV/mus) is enforced to reduce the system dead time. An estimate of the amplitude of such large signals is derived from the reset time, still obtaining a high resolution. Using this technique we achieved an energy resolution of 0.3% at 16.7 MeV