Published in

American Institute of Physics, Review of Scientific Instruments, 8(82), p. 084104

DOI: 10.1063/1.3628667

Links

Tools

Export citation

Search in Google Scholar

Infrared multiphoton dissociation tandem charge detection-mass spectrometry of single megadalton electrosprayed ions

This paper is available in a repository.
This paper is available in a repository.

Full text: Download

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Orange circle
Published version: archiving restricted
Data provided by SHERPA/RoMEO

Abstract

This work presents the implementation of tandem mass spectrometry for experiments on single electrosprayed ions from compounds of megadalton (MDa) molecular weight, using two charge detection devices. The first mass spectrometry stage (first charge detection device) combined with an ion gate allows both mass-to-charge ratio and charge selections of the megadalton ion of interest. The second stage is based on an electrostatic ion trap and consists of an image charge detection tube mounted between two ion mirrors. Single MDa ions can be stored for several dozen milliseconds. During the trapping time, single ions can be irradiated by a continuous wavelength CO(2) laser. We observe stepwise changes in the charge of a single trapped ion owing to multiphoton activation. Illustration of infrared multiphoton dissociation tandem mass spectrometry are given for single megadalton ions of poly(ethylene oxide)s and DNAs.