Elsevier, Procedia Engineering, (47), p. 1049-1052, 2012
DOI: 10.1016/j.proeng.2012.09.330
Full text: Download
It is proposed a microscopy for cell culture applications based on impedance sensors. The imagined signals are measured with the Electrical Cell-Substrate Spectroscopy (ECIS) technique, by identifying the cell area. The proposed microscopy allows real-time monitoring inside the incubator, reducing the contamination risk by human manipulation. It requires specific circuits for impedance measurements, a two-dimensional sensor array (pixels), and employing electrical models to decode efficiently the measured signals. Analogue Hardware Description Language (AHDL) circuits for cell-microelectrode enables the use of geometrical and technological data into the system design flow. A study case with 8x8 sensor array is reported, illustrating the evolution and power of the proposed image acquisition. (C) 2012 Elsevier Ltd....Selection and/or peer-review under responsibility of the Symposium Cracoviense Sp. z.o.o.