Published in

American Chemical Society, Analytical Chemistry, 18(84), p. 7599-7602, 2012

DOI: 10.1021/ac3019312

Links

Tools

Export citation

Search in Google Scholar

From Sample Processing to Quantification: A Full Electrochemical Approach for Neutral Analyte Derivatization, Capillary Electrophoresis Separation, and Contactless Conductivity Detection

This paper is available in a repository.
This paper is available in a repository.

Full text: Download

Green circle
Preprint: archiving allowed
  • Must obtain written permission from Editor
  • Must not violate ACS ethical Guidelines
Orange circle
Postprint: archiving restricted
  • Must obtain written permission from Editor
  • Must not violate ACS ethical Guidelines
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

A thin-layer electrochemical flow cell coupled to capillary electrophoresis with contactless conductivity detection (EC-CE-C(4)D) was applied for the first time to the derivatization and quantification of neutral species using aliphatic alcohols as model compounds. The simultaneous electrooxidation of four alcohols (ethanol, 1-propanol, 1-butanol, and 1-pentanol) to the corresponding carboxylates was carried out on a platinum working electrode in acid medium. The derivatization step required 1 min at 1.6 V vs. Ag/AgCl under stopped flow conditions, which was preceded by a 10 s activation at 0 V. The solution close to the electrode surface was then hydrodynamically injected into the capillary, and a 2.5 min electrophoretic separation was carried out. The fully automated flow system operated at a frequency of 12 analyses per hour. Simultaneous determination of the four alcohols presented detection limits of about 5 × 10(-5) mol L(-1). As a practical application with a complex matrix, ethanol concentrations were determined in diluted pale lager beer and in nonalcoholic beer. No statistically significant difference was observed between the EC-CE-C(4)D and gas chromatography with flame ionization detection (GC-FID) results for these samples. The derivatization efficiency remained constant over several hours of continuous operation with lager beer samples (n = 40).