Published in

European Geosciences Union, Atmospheric Measurement Techniques, 2(3), p. 387-395, 2010

DOI: 10.5194/amt-3-387-2010

European Geosciences Union, Atmospheric Measurement Techniques Discussions, 6(2), p. 3265-3290

DOI: 10.5194/amtd-2-3265-2009

Links

Tools

Export citation

Search in Google Scholar

First eddy covariance flux measurements by PTR-TOF

This paper is made freely available by the publisher.
This paper is made freely available by the publisher.

Full text: Download

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Green circle
Published version: archiving allowed
Data provided by SHERPA/RoMEO

Abstract

The recently developed PTR-TOF instrument was evaluated to measure methanol fluxes using the eddy covariance method. The PTR-TOF was employed at a well characterized temperate meadow field site in the Stubai valley, Austria, to measure VOC fluxes above an intensively farmed grass land. The high time resolution of the PTR-TOF allowed storing full mass spectra up to m/z 315 with a frequency of 10 Hz. Due to the high mass resolving power of the PTR-TOF three isobaric peaks were found at a nominal mass of m/z 33. Only one of the three peaks contributed to eddy covariance fluxes. The exact mass of this peak agrees well with the exact mass of protonated methanol (m/z 33.0335). The eddy covariance methanol fluxes measured with PTR-TOF were compared to virtual disjunct eddy covariance methanol fluxes simultaneously measured with a conventional PTR-MS. The methanol fluxes from both instruments show excellent agreement.