Published in

Institute of Electrical and Electronics Engineers, IEEE Transactions on Biomedical Engineering, 7(54), p. 1186-1190, 2007

DOI: 10.1109/tbme.2006.889193

Links

Tools

Export citation

Search in Google Scholar

Finite-Difference Modeling of the Anisotropic Electric Fields Generated by Stimulating Needles Used for Catheter Placement

This paper is available in a repository.
This paper is available in a repository.

Full text: Download

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

The use of peripheral nerve blocks to control pain is an increasing practice. Many techniques include the use of stimulating needles to locate the nerve of interest. Though success rates are generally high, difficulties still exist. In certain deeper nerve blocks, two needles of different geometries are used in the procedure. A smaller needle first locates a nerve bundle, and then is withdrawn in favor of a second, larger needle used for injection. The distinct geometries of these needles are shown to generate different electric field distributions, and these differences may be responsible for failures of the second needle to elicit nerve stimulation when placed in the same location as the first. A 3-D finite-difference method has been employed to numerically calculate the electric field distributions for a commercial set of stimulating needles.