Oxford University Press, Laboratory Medicine, 12(38), p. 723-725, 2007

DOI: 10.1309/95r8a56bdn10auuh

Links

Tools

Export citation

Search in Google Scholar

Evaluation of Different Mixing Procedures for K2 EDTA Primary Samples on Hematological Testing

This paper is made freely available by the publisher.
This paper is made freely available by the publisher.

Full text: Download

Green circle
Preprint: archiving allowed
Orange circle
Postprint: archiving restricted
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

Background: Although it is recommended that primary tubes containing an additive should be mixed several times, there is no evidence that unmixed specimens will provide unreliable results in hematological testing. Methods: Three primary 3.0-mL siliconized vacuum tubes containing 5.4 mg K2 ethylene diamine tetraacetic acid (EDTA) were sequentially collected from 20 healthy volunteers. The first was not inverted, was left standing in a vertical position, and then analyzed. The second and third tubes were respectively inverted 6 and 12 times immediately after collection and then analyzed. Results: When compared with the reference specimens inverted 6 times, results on unmixed specimens revealed significant decreases for red blood cell count, hemoglobin, hematocrit and platelets count, whereas the mean platelet volume was significantly increased. In none of the specimens were results flagged for platelet clumping, nor did the differences exceed the acceptable limits of bias. Conclusion: Provided the tubes are filled to their nominal volume, the blood-to-anticoagulant mixture occurring during sample collection may be adequate.