Published in

American Scientific Publishers, Journal of Biomedical Nanotechnology, 8(9), p. 1327-1335, 2013

DOI: 10.1166/jbn.2013.1619

Links

Tools

Export citation

Search in Google Scholar

Surface modification of stainless steel orthopedic implants by sol-gel ZrTiO4 and ZrTiO4-PMMA coatings

This paper is available in a repository.
This paper is available in a repository.

Full text: Download

Red circle
Preprint: archiving forbidden
Red circle
Postprint: archiving forbidden
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

In this paper, the biocompatibility of a medical-grade stainless steel coated with sol-gel derived, nanostructured inorganic ZrTiO4 and hybrid ZrTiO4-PMMA thin films is correlated with surface characteristics. The surfaces of the samples are characterized by atomic force microscopy, the sessile drop technique, and electrochemical corrosion experiments. The viability of adult human mesenchymal stem cells on the surfaces after one day of culture is also assessed quantitatively and morphologically. According to the results, both of the coatings improve the hydrophilicity, corrosion resistance, and thereby cytocompatibility of the substrate. Despite the higher corrosion protection by the hybrid coating, the sample coated with the inorganic thin film exhibits a better cell response, suggesting the domination of wettability. In summary, the ZrTiO4-based sol-gel films can be considered to improve the biocompatibility of metallic implants.