Magnetic induction tomography: hardware for multi-frequency measurements in biological tissues

Journal article published in 2001 by Hermann Scharfetter, Helmut K. Lackner ORCID, Javier Rosell
This paper is available in a repository.
This paper is available in a repository.

Full text: Download

Red circle
Preprint: archiving forbidden
Red circle
Postprint: archiving forbidden
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

Magnetic induction tomography (MIT) is a contactless method for mapping the electrical conductivity of tissue. MIT is based on the perturbation of an alternating magnetic field by a conducting object. The perturbation is detected by a voltage change in a receivercoil. At physiologically interesting frequencies (10 kHz-10 MHz) and conductivities (< 2 S m(-1)) the lower limit for the relative voltage change (signal/carrier ratio = SCR) to be resolved is 10(-7)-10(-10). A new MIT hardware has been developed consisting of a coil system with planar gradiometers and a high-resolution phase detector (PD). The gradiometer together with the PD resolves an SCR of 2.5 x 10(-5) (SNR = 20 dB at 150 kHz, acquisition speed: 100 ms). The system operates between 20 and 370 kHz with the possibility of extending the range up to 1 MHz. The feasibility of measuring conductivity spectra in the beta-dispersion range of biological tissues is experimentally demonstrated. An improvement of the resolution towards SCR = 10(-7) with an SNR of > or = 20 dB at frequencies > 100 kHz is possible. On-line spectroscopy of tissue conductivity with low spatial resolution appears feasible, thus enabling applications such as non-invasive monitoring of brain oedema.