Published in

Wiley, Journal of Magnetic Resonance Imaging, 1(9), p. 26-38, 1999

DOI: 10.1002/(sici)1522-2586(199901)9:1<26::aid-jmri4>3.0.co;2-h

Links

Tools

Export citation

Search in Google Scholar

Magnetic resonance imaging and cochlear implants: Compatibility and safety aspects

This paper is made freely available by the publisher.
This paper is made freely available by the publisher.

Full text: Download

Green circle
Preprint: archiving allowed
Orange circle
Postprint: archiving restricted
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

With cochlear implants, magnetic resonance imaging (MRI) has until recently been contraindicated due to excessive magnetic and electromagnetic interference. The aim of this study was to determine the MRI compatibility of the Med-E1 Combi 40/40+ cochlear implant, within a wide range of clinical MRI applications. In vitro experiments on a 1.5 T MR scanner were performed. Torque, force, demagnetization, artifacts, induced voltages, and temperature increase were measured in worst case scenarios for the implant. In addition, stabilization experiments were performed. It was shown that most of the electromagnetic interferences between the cochlear implant and the 1.5 T scanner remained within acceptable limits. One exception is the torque on the internal magnet, which represents a hazard for patients with these cochlear implants. Therefore, MRI examination should only be performed if there is a strong medical indication, and certainly some assessment of the relative risks involved versus the risk of not providing the diagnostic capabilities of MRI, will have to be made. Appropriate safety measures should be taken.