Published in

Elsevier, Physiology & Behavior, (118), p. 185-188

DOI: 10.1016/j.physbeh.2013.05.036

Links

Tools

Export citation

Search in Google Scholar

Polysomnographic validation of a wireless dry headband technology for sleep monitoring in healthy young adults

This paper is available in a repository.
This paper is available in a repository.

Full text: Download

Green circle
Preprint: archiving allowed
Orange circle
Postprint: archiving restricted
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

Abstract The present study aimed to explore the validity and reliability of a wireless dry headband technology for sleep monitoring (WS), through a comparison with concurrent polysomnographic (PSG) recording in healthy young adults. Eleven volunteers (7 females; mean age ± SD: 24.75 ± 3.62 years) took part in the study, wearing the WS for two overnight PSG recordings in the sleep laboratory. The WS was compared to PSG in the identification of wake, light, deep and REM sleep. The WS sensitivity and specificity were 97.6% and 56.1%, respectively. The WS agreement with PSG, measured by Cohen’s kappa, was 0.56 for light sleep, 0.70 for deep sleep, and 0.67 for REM sleep. The present results showed that the agreement ranged from moderate to high between PSG and the WS, while wakefulness detection was observed to be a limitation of the WS.