Geometric accuracy evaluation of the new VERO stereotactic body radiation therapy system

This paper is made freely available by the publisher.
This paper is made freely available by the publisher.

Full text: Download

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Green circle
Published version: archiving allowed
Data provided by SHERPA/RoMEO

Abstract

Real-time tracking of moving tumors is one of today's challenges in radiation therapy. This work investigates the tracking performance of VERO, a novel treatment device with gimbaled linear accelerator, especially designed for four dimensional image guided radiotherapy. It is found that the significant impact of organ motion on dose distribution can be overcome by combining a polynomial predictor with a prediction horizon of 50ms with the VERO tracking system. Tracking errors can be reduced from 1.7mm to 0.6mm for realistic patient signals as well as sine wave from 5 to 30 bpm.