Published in

Elsevier, Talanta, 5(44), p. 867-875

DOI: 10.1016/s0039-9140(96)02131-5

Links

Tools

Export citation

Search in Google Scholar

Flow injection determination of Pb and Cd traces with graphite furnace atomic absorption spectrometry

Journal article published in 1997 by O. Abollino, M. Colognesi, M. Aceto ORCID, C. Sarzanini, E. Mentasti
This paper was not found in any repository, but could be made available legally by the author.
This paper was not found in any repository, but could be made available legally by the author.

Full text: Unavailable

Green circle
Preprint: archiving allowed
Red circle
Postprint: archiving forbidden
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

The preconcentration and recovery of lead and cadmium traces at ng l(-1) level were evaluated in standard solutions and natural aqueous samples using a FIAS (Flow Injection Atomic Spectrometry) apparatus. The method is based on retention of the complex formed between Pb or Cd and 1,2-dihydroxy-3,5-benzendisulphonic acid (Tiron) on a macroporous anion-exchange resin. The recovery of the analytes was obtained by elution with 0.1 M HCl and their determination was performed by Graphite Furnace Atomic Absorption Spectrometry (GFAAS). The detection limits were 9 and 7 ng l(-1) for Pb and Cd respectively. The effects of sample solution pH and composition and of interfering agents as well as reagent purity are discussed. The technique was applied to the analysis of natural waters.