Full text: Download
PURPOSE: X-ray mammography is the primary tool for early detection of breast cancer and for follow-up after breast conserving therapy (BCT). BCT-treated breasts are smaller, less elastic, and more sensitive to pain. Instead of the current force-controlled approach of applying the same force to each breast, pressure-controlled protocols aim to improve standardization in terms of physiology by taking breast contact area and inelasticity into account. The purpose of this study is to estimate the potential for pressure protocols to reduce discomfort and pain, particularly the number of severe pain complaints for BCT-treated breasts. METHODS: A prospective observational study including 58 women having one BCT-treated breast and one untreated nonsymptomatic breast, following our hospital's 18 decanewton (daN) compression protocol was performed. Breast thickness, applied force, contact area, mean pressure, breast volume, and inelasticity (mean E-modulus) were statistically compared between the within-women breast pairs, and data were used as predictors for severe pain, i.e., scores 7 and higher on an 11-point Numerical Rating Scale. Curve-fitting models were used to estimate how pressure-controlled protocols affect breast thickness, compression force, and pain experience. RESULTS: BCT-treated breasts had on average 27% smaller contact areas, 30% lower elasticity, and 30% higher pain scores than untreated breasts (allp