Dissemin is shutting down on January 1st, 2025

Published in

2010 Annual International Conference of the IEEE Engineering in Medicine and Biology

DOI: 10.1109/iembs.2010.5626132

Links

Tools

Export citation

Search in Google Scholar

Evaluation of a CE approved ambulatory patient monitoring device in a general medical ward

This paper is available in a repository.
This paper is available in a repository.

Full text: Download

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

An evaluation of a newly CE approved bedside monitoring device used in a general hospital ward is presented. This evaluation has shown that it is feasible to use the system within this environment to provide medical staff with supplementary information on patient health, at more frequent intervals than traditional monitoring methods. The physiological data recorded by the body worn device is wirelessly transmitted to a patient management system for storage and display. Good correlation between heart rate values recorded by hospital staff and those recorded by the automated Vitalsens VS100 system was observed. The system has highlighted clinical information that routine observations alone did not readily identify. This can provide clinicians with a better view of the overall health status of the patient. Such medical issues include those witnessed in this study, namely paroxysmal AF, ectopic beats, increasing heart rates recorded prior to a hypoglycaemic event, general high and low heart rate trends and various instances where clinically relevant ECG data has been captured.