Elsevier, Academic Radiology, 3(8), p. 219-224
DOI: 10.1016/s1076-6332(03)80530-5
Full text: Unavailable
RATIONALE AND OBJECTIVES: The purpose of this study was to comparatively evaluate digital planar mammography and both linear and nonlinear tomosynthetic reconstruction methods. MATERIALS AND METHODS: A "disk" (ie, target) identification study was conducted to compare planar and reconstruction methods. Projective data using a composite phantom with circular disks were acquired in both planar and tomographic modes by using a full-field, digital mammographic system. Two-dimensional projections were reconstructed with both linear (ie, backprojection) and nonlinear (ie, maximization and minimization) tuned-aperture computed tomographic (TACT) methods to produce three-dimensional data sets. Four board-certified radiologists and one 4th-year radiology resident participated as observers. All images were compared by these observers in terms of the number of disks identified. RESULTS: Significant differences (P