Multi-Probe RFA vs. Single-Probe MWA in an Ex Vivo Bovine Liver Model: Comparison of Volume and Shape of Coagulation Zones

Journal article published in 2023 by Yannick Scharll, Peter Schullian, Michael Bauer, Gregor Laimer ORCID, Reto Bale
This paper is made freely available by the publisher.
This paper is made freely available by the publisher.

Full text: Download

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Green circle
Published version: archiving allowed
Data provided by SHERPA/RoMEO

Abstract

Objectives: To compare the volumes and shapes of the coagulation zone (CZ) of a multi-probe RFA system (three RFA electrodes) and a single-probe MWA system from the same vendor in an ex vivo bovine liver model. Material & Methods: A total of 48 CZs were obtained in bovine liver specimens with three different ablation system configurations (single-probe MWA vs. multi-probe RFA with 20 mm inter-probe distance [confluent CZ] vs. multi-probe RFA with 50 mm inter-probe distance [three individual CZs]) at 4, 6, 8, and 10 min ablation time using a fixed ablation protocol. Ablation diameters were measured and ellipticity indices (EIs) and volumes calculated. Calculations for all systems/configurations were compared. Results: Volumes and diameters increased with ablation time for all configurations. At 4 and 6 min ablation time volumes obtained with the RFA 50 mm setup, and at 8 and 10 min with the RFA 20 mm setup were the largest at 26.5 ± 4.1 mL, 38.1 ± 5.8 mL, 46.3 ± 4.9 mL, 48.4 ± 7.3 mL, respectively. The single-probe MWA could not reach the volumes of the RFA setups for any of the ablation times evaluated. EI were very similar and almost round for RFA 20 mm and single-probe MWA, and differed significantly to the more ovoid ones for the RFA 50 mm configuration. Conclusions: The multi-probe RFA system employing three electrodes achieved significantly larger ablation volumes in both configurations (confluent CZ and three individual CZs) per time as compared with a single-probe MWA system in this ex vivo bovine liver model.