Design and Optimization of a Wearable Underactuated Mechanism for Spinal Posture Measurement

Journal article published in 2024 by Ming-Chang Hsu, Hsuan-Yu Chen ORCID, Christina Soong, T.-J. Yeh
This paper was not found in any repository, but could be made available legally by the author.
This paper was not found in any repository, but could be made available legally by the author.

Full text: Unavailable

Green circle
Preprint: archiving allowed
Green circle
Postprint: archiving allowed
Red circle
Published version: archiving forbidden
Data provided by SHERPA/RoMEO

Abstract

Abstract This paper proposes a novel wearable device to monitor and record the posture and alignment of spine. The proposed device adopts an underactuated mechanism design which allows it to adapt to the multiple-degrees-of-freedom spinal posture with minimum weight and complexity. To ensure the validity of measurement and comfort of wearing, the mechanism parameters are determined first by considering a special posture and then are fine-tuned using an optimization algorithm so that uniform contact forces for several selected spinal postures can be achieved. Experiments demonstrate that the device can automatically maintain contact with the wearer’s back and offer real-time spinal posture and alignment data for medical diagnosis and treatment.